• Title of article

    The measurement of activity contained in a 32P stainless-steel stent by destructive assay

  • Author/Authors

    Jeffrey T. Cessna، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    6
  • From page
    337
  • To page
    342
  • Abstract
    Destructive assays of stainless-steel stents containing 32P were performed. Prior to assay, 24 stents were intercompared on a NaI(Tl) well-type crystal. A subset of stents were then digested in a suitable carrier solution by the addition of concentrated hydrofluoric acid. The final solution was measured by the CIEMAT/NIST tritium efficiency tracing method for liquid scintillation counting. A separate experiment was performed which indicated no loss of activity during stent digestion. Expanded (k=2) uncertainties on stent activity ranged from 1.0% to 2.6%.
  • Keywords
    Liquid-scintillation , restenosis , 32P , Radioactivity measurement , angioplasty