Title of article
A rapid improved method for gamma-spectrometric determination of 202Tl impurities in [201Tl]-labelled radiopharmaceuticals
Author/Authors
Mauro Bonardi، نويسنده , , Flavia Groppi، نويسنده , , Claudio Birattari، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2002
Pages
9
From page
647
To page
655
Abstract
Despite the cyclotron production method and the efficiency of the radiochemical procedures adopted, the long-lived radio-isotopic impurity 202Tl is always present in [201Tl]-labelled radio-pharmaceuticals, together with other short-lived impurities like, 200Tl. Rapid determination of the 202Tl impurity, can be achieved using HPGe gamma spectrometry and a detector shielded by a 5 mm thick envelope of lead. In this way, dead-time correction errors, Compton and X-ray background, are very efficiently avoided and suppressed. The same method could be applied routinely in nuclear medicine, to determine the radioisotopic purity of 201Tl by means of an ionisation chamber dose calibrator.
Keywords
Absorption , g-spectrometry , Dead time , Sample shielding , 202Tl impurity , 201Tl
Journal title
Applied Radiation and Isotopes
Serial Year
2002
Journal title
Applied Radiation and Isotopes
Record number
541349