Title of article
An investigation of the repeatability of calibration factors in gamma-ray spectrometry of geological materials
Author/Authors
A. O. Mustapha، نويسنده , , J. P. Patel، نويسنده , , I. V. S. Rathore، نويسنده , , N. O. Hashim، نويسنده , , D. Otwoma، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
4
From page
79
To page
82
Abstract
A NaI(Tl)-based gamma spectrometer for the analysis of geological materials was calibrated using the IAEA reference materials RGU-1, RGTH-1 and RGK-1. To simulate typical geological samples, two additional standards were prepared from aliquots of the three reference materials. The reproducibility of the instrument calibration factors (CFs) was tested by repeated measurements of the pure IAEA reference materials and the mixed samples in a reproducible counting geometry. The results were analysed using a two-way classification analysis of variance; it was found that the variance in the CFs is significantly higher between standards than it is between measurements. Allowance should be made for this when estimating uncertainties in measurements with the NaI(Tl) spectrometers.
Keywords
calibration , Gamma spectrometer , IAEA reference materials , naturally occurring radionuclides , Analysis of variance
Journal title
Applied Radiation and Isotopes
Serial Year
2004
Journal title
Applied Radiation and Isotopes
Record number
541562