• Title of article

    A tunable monochromatic X-ray source for metrological studies in the 1–20 keV energy range: application to the measurement of attenuation coefficients

  • Author/Authors

    Marie-Christine Lépy، نويسنده , , Laurent Ferreux، نويسنده , , Johann Plagnard، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    7
  • From page
    159
  • To page
    165
  • Abstract
    A tunable monochromatic X-ray source operating in the 1–20 keV energy range is described. An X-ray tube provides initial photons. A dispersive crystal performs the energy selection, according to Braggʹs law. An X-ray detector is connected to the monochromator fixed exit. This setup can be used for metrological studies. A first application consists in measuring attenuation coefficients in the 4–10 keV energy range. Results for aluminum and copper are given, with average relative uncertainties (1σ) of 1% and 3% respectively.
  • Keywords
    x-ray source , Monochromator , attenuation coefficient , aluminum , copper , Detector calibration
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2004
  • Journal title
    Applied Radiation and Isotopes
  • Record number