Title of article
Application of GESPECOR software for the calculation of coincidence summing effects in special cases
Author/Authors
Dirk Arnold، نويسنده , , Octavian Sima، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
6
From page
167
To page
172
Abstract
In this work, coincidence summing correction factors have been measured for 133Ba, 152Eu and 88Y point sources with a 50% relative efficiency p-type detector and a 25% relative efficiency n-type detector in two close-to-detector measurement geometries. The experimental data for 133Ba and 152Eu and the results obtained with the GESPECOR software reveal a complex structure of the conventional dead layer of the p-type detector. The high value of the coincidence summing correction factor for the 511 keV peak of 88Y, in agreement with the values computed by GESPECOR, in this case cautions against the application of the semiempirical method for evaluating coincidence summing effects.
Keywords
Gamma-ray spectrometry , Coincidence summing effects , Detector dead layer , Ge detectors
Journal title
Applied Radiation and Isotopes
Serial Year
2004
Journal title
Applied Radiation and Isotopes
Record number
541579
Link To Document