Title of article
Measurements of nuclear data parameters of 201Tl by gamma-ray spectrometry
Author/Authors
Karla C. de Souza، نويسنده , , Mônica L. da Silva، نويسنده , , José U. Delgado، نويسنده , , Roberto Poledna، نويسنده , , Ricardo T. Lopes، نويسنده , , Carlos J. da Silva، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
4
From page
307
To page
310
Abstract
201Tl is frequently used in radiopharmaceutical applications, and therefore the gamma-ray emission probabilities and half-life have been re-determined by means of gamma-ray spectroscopy. While the activity was calibrated using the sum-peak coincidence method, the half-life was obtained by the reference source method based on simultaneous counting of a reference source and the sample. Both the measurement techniques and assignment of uncertainties are presented and discussed, and the resulting data are shown to be in good agreement with previously published studies.
Keywords
Gamma-ray spectrometry , Half-life determination , Gamma-ray emission probability , Sum-peak method