• Title of article

    Absolute radioactivity measurements by the use of a 4πβ–4πγ detector configuration

  • Author/Authors

    Yasushi Kawada، نويسنده , , Masashi Ohtuka، نويسنده , , Qiu-Wei Wang، نويسنده , , Yoshio Hino، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    357
  • To page
    362
  • Abstract
    The radioactivity of 60Co and 134Cs sources were measured using a 4πβ–4πγ detector configuration with a well-type NaI(Tl) crystal and a sandwich type 4πβ detector composed of two sheets of NE102A plastic scintillator coupled to a slender photomultiplier tube. The β-detector was inserted into the well of the γ-detector. Since counting efficiencies in both the β- and γ-channels can be kept high, excellent counting statistics were attainable even when weak sources were measured. This configuration can also be used for radioactivity measurements based upon the direct integral counting of 4πβ+4πγ logic sum signals, and two independent results can be obtained for every measurement. This technique is especially useful for the standardization of complex decaying nuclides.
  • Keywords
    Coincidence counting , Well-type detector , Integral counting , Plastic sandwich scintillator , radioactivity , 4??–4??
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2004
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    541610