Title of article
Low-level gamma spectrometry using beta coincidence and Compton suppression
Author/Authors
E. L. Grigorescu، نويسنده , , P. de Felice، نويسنده , , Anamaria Cristina Razdolescu، نويسنده , , A. Luca، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
5
From page
191
To page
195
Abstract
A low-level gamma-ray spectrometry system was developed using a Ge(Li) detector with 6% relative efficiency coupled to a 2π beta plastic detector for coincidence selection and a massive NaI(Tl) detector for Compton suppression.
The integral background count rate for (50–1500) keV was 0.5 s−1 kg−1 (Ge), using only beta coincidences. With Compton suppression, a value of 0.25 s−1 kg−1 (Ge) was obtained. Spectra with and without Compton suppression were studied for 60Co, 137Cs and 152Eu point sources. Considerations are made concerning the Compton suppression advantages in different situations.
Keywords
Low-level gamma spectrometry , Detection limit , Compton suppression
Journal title
Applied Radiation and Isotopes