• Title of article

    Dosimetric comparison at FiR 1 using microdosimetry, ionisation chambers and computer simulation

  • Author/Authors

    J. Uusi-Simola، نويسنده , , T. Serén، نويسنده , , T. Sepp?l?، نويسنده , , A. Kosunen، نويسنده , , I. Auterinen، نويسنده , , S. Savolainen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    4
  • From page
    845
  • To page
    848
  • Abstract
    Tissue equivalent proportional counter microdosimetry has been applied in the dosimetry of epithermal neutron beams as they can provide an independent and accurate method to determine gamma ray and neutron absorbed doses. Dosimetric comparison has been performed using a tissue equivalent proportional counter, dual ionisation chambers and DORT computer code at FiR 1 boron neutron capture therapy facility in Espoo, Finland. The three methods were applied to determine neutron and gamma ray absorbed doses at 25, 40, 60 and 120 mm depths along the beam centerline in a water-filled PMMA phantom. The determined absorbed doses were found to agree within the limits of the estimated uncertainties.
  • Keywords
    microdosimetry , epithermal neutron beam , BNCT
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2004
  • Journal title
    Applied Radiation and Isotopes
  • Record number