• Title of article

    Theoretical estimates of the solid angle subtended by a dual diaphragm–detector assembly for alpha sources

  • Author/Authors

    Julio C. Aguiar، نويسنده , , Eduardo Galiano، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    3
  • From page
    1349
  • To page
    1351
  • Abstract
    Knowledge of the solid angle subtended by a detector is essential for the determination of the activity of a radioactive source. This work investigates the influence of diaphragm and source diameter, and source to diaphragm distance, on the solid angle subtended by a dual opposed α diaphragm–detector assembly. Expressions published by Segre, Ruby, and others, and the Monte Carlo method are used. The analytical expressions consistently produced higher estimates than the Monte Carlo method for the solid angle.
  • Keywords
    Radiation detector , Monte Carlo method , Geometric efficiency , Solid angle
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2004
  • Journal title
    Applied Radiation and Isotopes
  • Record number