• Title of article

    Compensation by the CIEMAT/NIST-method of long-term effects in LSC measurements of beta emitters

  • Author/Authors

    L. Rodriguez Barquero، نويسنده , , J. M. Los Arcos، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    9
  • From page
    1403
  • To page
    1411
  • Abstract
    A systematic evaluation of activity concentrations of 3H, 63Ni and 14C by using the CIEMAT/NIST method has been carried out over time intervals of 4, 2.4 and 2.3 years, respectively. It proves that the CIEMAT/NIST method can compensate, with no need for additional checking sources or procedures, for count losses and quench parameter reduction in long-term measurements with LSC spectrometers and allows one to obtain very accurate activity concentration values by using a quench curve close to the nuclide measurements. If this requirement of the method is not fulfilled, significant discrepancies, up to 5.6% for 3H, 1.5% for 63Ni and 0.35% for 14C, can be obtained in only 1-year time difference.
  • Keywords
    CIEMAT/NIST method , Beta emitters , Liquid scintillation counting , Long-term stability