• Title of article

    Evaluation of the depth-ratio and backscattering-factor methods in quality control measurements of electron beam energies

  • Author/Authors

    Cleber N. De Souza، نويسنده , , Cl?udio H. Sibata، نويسنده , , Linda V.E Caldas، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    217
  • To page
    222
  • Abstract
    Sensitivities of two different methods for evaluating electron beam energy have been investigated. One of them uses the ratio of ionization at two depths corresponding to two different points on the depth curve of ionization. The other one uses the magnitude of backscattering by lead as a measure of the electron beam energy. The depth-ratio method was found to be sensitive and practical for a routine quality control program. The sensitivity of the backscattering effect was found to be insufficient for electron energies higher than 12 MeV.
  • Keywords
    electron beam , energy measurement , quality assurance , electron backscattering
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2005
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    542098