• Title of article

    Determination of the activity concentration of a 238Pu solution by the defined solid angle method utilizing a novel dual diaphragm-detector assembly

  • Author/Authors

    Julio C. Aguiar، نويسنده , , Eduardo Galiano، نويسنده , , Pablo Arenillas، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    229
  • To page
    233
  • Abstract
    The activity concentration of a 238Pu solution was measured by the determined solid angle method employing a novel dual diaphragm-detector assembly, which has been previously described. Due to the special requirements of the detector, a new type of source holder was developed, which consisted of sandwiching the radioisotope between two organic films called VYNS. It was experimentally demonstrated that the VYNS films do not absorb α particles, but reduce their energy by an average of 22 keV.A mean activity concentration for 238Pu of 359.10±0.8 kBq/g was measured.
  • Keywords
    238Pu activity concentration , Dual ? particle detector , VYNS film , VYNS absorption , VYNS attenuation
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2005
  • Journal title
    Applied Radiation and Isotopes
  • Record number