    In vitro, in vivo and post explantation testing of glucose-detecting biosensors: Current methods and recommendations

  • Author/Authors

    Heidi E. Koschwanez، نويسنده , , William M. Reichert، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    17
  • From page
    3687
  • To page
    3703
  • Abstract
    To date, there have been a number of cases where glucose sensors have performed well over long periods of implantation; however, it remains difficult to predict whether a given sensor will perform reliably, will exhibit gradual degradation of performance, or will fail outright soon after implantation. Typically, the literature emphasizes the sensor that performed well, while only briefly (if at all) mentioning the failed devices. This leaves open the question of whether current sensor designs are adequate for the hostile in vivo environment, and whether these sensors have been assessed by the proper regimen of testing protocols. This paper reviews the current in vitro and in vivo testing procedures used to evaluate the functionality and biocompatibility of implantable glucose sensors. An overview of the standards and regulatory bodies that govern biomaterials and end product device testing precedes a discussion of up-to-date invasive and non-invasive technologies for diabetes management. Analysis of current in vitro, in vivo, and then post explantation testing is presented. Given the underlying assumption that the success of the sensor in vitro foreshadows the long-term reliability of the sensor in the human body, the relative merits of these testing methods are evaluated with respect to how representative they are of human models.
  • Keywords
    Biosensor , Foreign body response , in vitro test , wound healing , In vivo test
  • Journal title
    Biomaterials
  • Serial Year
    2007
  • Journal title
    Biomaterials
  • Record number

    547651