Close-geometry efficiency calibration of p-type HPGe detectors with a Cs-134 point source
Author/Authors
P. DeFelice، نويسنده , , A. Fazio، نويسنده , , T. Vidmar، نويسنده , , M. Korun، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
4
From page
1303
To page
1306
Abstract
When close-geometry detector calibration is required in gamma-ray spectrometry, single-line emitters are usually used in order to avoid true coincidence summing effects. We managed to overcome this limitation by developing a method for the determination of the efficiency of p-type HPGe detectors in close-geometry with a calibrated Cs-134 point source. No separate determination of coincidence summing correction factors is required and a single measurement furnishes the full-energy-peak efficiencies in the 475–1365 keV energy range.
Keywords
True coincidence summing , Close geometry , Detector calibration , Gamma-ray spectrometry , Cs-134