Title of article
Experimental and MC determination of HPGe detector efficiency in the 40–2754 keV energy range for measuring point source geometry with the source-to-detector distance of 25 cm
Author/Authors
Pavel Dryak، نويسنده , , Petr Kovar، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
4
From page
1346
To page
1349
Abstract
A precise model of a 40% relative efficiency p-type HPGe detector was created for photon detection efficiency calculation using the MCNP code. All detector parameters were determined by different experiments. No experimental calibration points were used for the modification of detector parameters. The model was validated by comparing calculated and experimental full energy peak efficiencies in the 40–2754 keV energy range, for point-source geometry with the source-to-detector distance of 25 cm.