Title of article
Absolute counting of 188Re radiopharmaceuticals
Author/Authors
Ming-Chen Yuan، نويسنده , , Hsiao-Fang Pang، نويسنده , , Chu-Fang Wang، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
4
From page
1380
To page
1383
Abstract
The Institute of Nuclear Energy Research (INER) has provided 188Re radiopharmaceuticals for hospitals in Taiwan. To enhance the accuracy of commercial radionuclide calibrators used by radiopharmacies and hospitals, and to ensure that patients receive proper doses of these radiopharmaceuticals, it is very important to standardize this nuclide. The 4πβ-γ coincidence counting method was used to standardize the mass activities of 188Re in this study. At the same time, three well type ionization chambers, Centronic IG11-N20, Centronic IG11-A20, and ISOCAL-IV, were calibrated by the standardized solutions of the nuclide. In this research, the calibration figures of ISOCAL-IV for the nuclide were consistent with the results of National Physical Laboratory (NPL, UK). The outcome implied that the results of coincidence counting did mutually agree. On the other hand, the radionuclide calibrator in the radiopharmacy was producing measurement errors of about 20% when using the manufacturers recommended calibration setting: an accurate correction factor has now been determined in this study.
Keywords
188Re , Standardization , Coincidence counting
Journal title
Applied Radiation and Isotopes
Serial Year
2006
Journal title
Applied Radiation and Isotopes
Record number
548083
Link To Document