• Title of article

    Calibration of a 137Cs γ-ray beam irradiator using large size chambers

  • Author/Authors

    R. Minniti، نويسنده , , S.M. Seltzer، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    6
  • From page
    401
  • To page
    406
  • Abstract
    A 137Cs γ-ray beam irradiator has been calibrated in terms of air kerma using large size chambers. The available air-kerma rates range between 1.8 μGy/h (0.2 mR/h) and 5.3 mGy/h (0.6 R/h). Large-volume chambers were used to characterize the source in terms of the radiation quantity air kerma (and exposure). Two types of chambers with significantly different characteristics and energy responses were used. This work shows that very good agreement can be obtained between the measurements performed with such different types of chambers. An agreement of 0.3% is observed between chambers even for the lowest air-kerma rates measured.
  • Keywords
    Calibration coefficient , ionization chamber , Cs-137 , ?-Ray , Secondary standard , Scattered radiation , Air kerma
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2007
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    548186