• Title of article

    Investigation of photoneutron dose equivalent from high-energy photons in radiotherapy

  • Author/Authors

    Jao-Perng Lin، نويسنده , , Wei-Chung Liu، نويسنده , , Chun-Chih Lin، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    6
  • From page
    599
  • To page
    604
  • Abstract
    Spatial distribution of photoneutron dose equivalent during radiotherapy at different beam size, depth, and distance from a 15 MV linear accelerator was investigated with bubble detectors in a water phantom. The photoneutron dose equivalent was mainly from fast neutrons, and decreased with distance at a fixed field and with depth. Besides, photoneutron dose equivalent was slightly affected by beam size due to the variation of tungsten area exposed in the beam direction and photoneutrons occurred at the jaws. Fast photoneutron dose equivalent of shallow critical organs was represented still considerably outside the beam size.
  • Keywords
    accelerator , Bubble detector , Neutron , dose equivalent , Photoneutron , radiotherapy
  • Journal title
    Applied Radiation and Isotopes
  • Serial Year
    2007
  • Journal title
    Applied Radiation and Isotopes
  • Record number

    548216