Title of article
X-Ray energy dependence of the dose response of SIRAD radiation dosimeters
Author/Authors
Tsang Cheung، نويسنده , , Martin J. Butson، نويسنده , , Peter K.N. Yu، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
4
From page
814
To page
817
Abstract
SIRADs (self-indicating instant radiation alert dosimeters) are designed to measure accident radiation doses. As the energy of radiation is usually unknown in such situations, a detector with a weak energy dependence of its response to dose would be ideal. We have studied the energy dependence of the dose response of SIRADs in the range from 50 kVp to 10 MV, which corresponds to photon equivalent energies from 25.5 keV to 2.2 MeV. The response to the same dose at 25.5 keV is (29±4)% (±1s) lower than the response at 1.4 MeV. The response to a dose slowly increases with radiation energy. This energy dependence is relatively weak in comparison with the dependence for radiographic films and similar in magnitude to the dependence for lithium fluoride thermoluminescence dosimeters. This energy dependence of the response diminishes the accuracy of dose assessments in radiation fields of unknown energy, but does not significantly compromise the core ability of the devices to provide visual estimates of radiation doses.
Keywords
SIRAD , Accident dosimetry , dose response , energy dependence , radiochromic film
Journal title
Applied Radiation and Isotopes
Serial Year
2007
Journal title
Applied Radiation and Isotopes
Record number
548248
Link To Document