Evaluation of different counting methods for use in radiochemical purity testing procedures for 99mTc-labelled radiopharmaceuticals
Author/Authors
Claudio Maioli، نويسنده , , Alberto Bestetti، نويسنده , , Franco Milani، نويسنده , , Gian Paolo Cornalba، نويسنده , , Luca Tagliabue، نويسنده , , Domenica Di Benedetto، نويسنده , , Ilaria Rognoni، نويسنده , , Gian Luigi Tarolo، نويسنده , , Rita Paroni، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
4
From page
556
To page
559
Abstract
The efficiency and accuracy of different methods for quality control of radiopharmaceutical preparations for diagnostic purpose were studied. The radiochemical purity of 99mTc Tetrafosmin, 99mTc Exametazime, 99mTc Sestamibi and 99mTc Oxidronate was evaluated by different thin layer chromatography systems, followed by cutting of the strips into two or three sections and by the measurement of radioactivity distribution by dose calibrator or gamma counter. In addition, to confirm the accuracy of these routine procedures, the strips were cut into a number of micro-sections (14–25) and each of them evaluated by the gamma counter.
The three tested procedures gave similar results and revealed a good and comparable accuracy. The radioactivity measurement with the dose calibrator remains the most practicable because of the rapidity of execution.
Keywords
Radiochemical purity , Radiopharmaceuticals , quality control , ITLC-SG chromatography