Title of article
A semi-automated system for concentration of rhenium-188 for radiopharmaceutical applications
Author/Authors
G. Wunderlich، نويسنده , , H. Hartmann، نويسنده , , M. Andreeff، نويسنده , , J. Kotzerke، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
5
From page
1876
To page
1880
Abstract
A semi-automated system has been designed and installed for the routine production of highly concentrated solutions of rhenium-188 perrhenate in saline which incorporates extensive radiation protection for the laboratory staff. This new system represents a simple, rapid, and effective computer controlled method using inexpensive disposable components. The flipper-valve settings and the radioactivity measured by use of collimated gamma dose rate probes are continuously monitored. If necessary, 2–3 tungsten-188/rhenium-188 generators can be eluted in a tandem arrangement. Eluate volumes (1.3–2 ml) of maximum specific radioactivity of 25 GBq/ml were routinely prepared.
Keywords
Specific activity , Nuclear medicine therapy , Beta-emitter , Re-generator , 188Re