• Title of article

    Perfusion imaging of skin island flap blood flow by a scanning laser-Doppler technique

  • Author/Authors

    F. Arnold، نويسنده , , C.F. He، نويسنده , , C.Y. Jia، نويسنده , , G.W. Cherry، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    8
  • From page
    280
  • To page
    287
  • Abstract
    Conventional laser-Doppler perfusion measurements can only obtain information from a single site. Since superficial blood flow is heterogeneous, this is a serious limitation, particularly in studies of methods to improve skin flap survival. A scanning laser-Doppler instrument has been developed which provides both an image and quantitative information about perfusion of the superficial tissue. We have evaluated this instrument in a circumflex iliac artery island flap model in the pig. The validity of the model was demonstrated by fluorescein dye injection and histology. Elevation of flaps was found to increase proximal flap blood flow but to decrease it in the distal portion. In flaps subjected to arterial ischaemia (9 h) and reperfusion we found an early increase in proximal flow, which gradually extended distally over the first 40 to 60 min. 16 h later, flow had declined, compatible with reperfusion damage to the vasculature.
  • Journal title
    Journal of Plastic , Reconstructive and Aesthetic Surgery
  • Serial Year
    1995
  • Journal title
    Journal of Plastic , Reconstructive and Aesthetic Surgery
  • Record number

    593935