Title of article
Novel surgical system for reducing lung tissue and preventing air leaks
Author/Authors
Eric Vallières، نويسنده , , Xavier Gonzalez، نويسنده , , Kimberly M Pedersen، نويسنده , , Gena K Sears، نويسنده , , Steven C Springmeyer، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
4
From page
2071
To page
2074
Abstract
Purpose
This study assessed the feasibility of using the VALR surgical system (Spiration Inc, Redmond, WA), limited by federal law to investigational use, for capturing and reducing a selected portion of affected lobes in patients undergoing lobectomy.
Description
The tested system consists of a hand-held vacuum-regulated introducer loaded with a flexible, silicone sleeve. Targeted tissue is drawn into the introducer and the silicone sleeve is deployed and sutured in place. The end of the proximal sleeve includes a compression band for applying uniform radial pressure, suture ports, and silicone lugs lining the inner lumen for reinforcing sleeve position.
Evaluation
The system was effective in capturing 25% to 30% tissue of each lobe tested. Mean intraoperative test time was 8.5 minutes. The compression sleeve did not slip or dislodge after suturing, and no tissue damage or leaks were observed.
Conclusions
It was feasible using vacuum to draw and isolate a portion of pulmonary tissue within a silicone sleeve. The system was intuitive to apply, easy to use, and produced effective reduction and sealing of tissue.
Journal title
The Annals of Thoracic Surgery
Serial Year
2003
Journal title
The Annals of Thoracic Surgery
Record number
607144
Link To Document