• Title of article

    Portable Coronary Active Perfusion System for Off-Pump Coronary Artery Bypass Grafting

  • Author/Authors

    Yoshinao Koshida، نويسنده , , Go Watanabe، نويسنده , , Tamotsu Yasuda، نويسنده , , Shigeyuki Tomita، نويسنده , , Shinichi Kadoya، نويسنده , , Taro Kanamori، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    706
  • To page
    710
  • Abstract
    pose The present study was performed to develop a new perfusion system for off-pump coronary artery bypass grafting and to examine whether even a simple coronary perfusion system can maintain adequate blood flow delivery during anastomosis. Description The experiment was performed in two stages. In procedure 1, 3 pigs with left anterior descending artery occlusion were used to evaluate optimal perfusion flow rate and coronary artery internal pressure, and to evaluate the safety area of perfusion. In procedure 2, 6 pigs were used to validate the new portable coronary perfusion system. Evaluation The optimal blood flow in the portable coronary active perfusion system was less than approximately 40 mL/min. The small, easy to use pump system (ie, the portable coronary active perfusion system) may prevent hemodynamic deterioration and ventricular arrhythmia during coronary occlusion, resulting in better maintenance of left ventricular function. Conclusions Even a simple pump system can achieve effective perfusion for safe anastomosis. Further studies are required to allow the clinical use of this system.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    2006
  • Journal title
    The Annals of Thoracic Surgery
  • Record number