• Title of article

    Late Disruption of a Freestyle Stentless Bioprosthesis Used for Repair of Sinus of Valsalva Aneurysm of Noncoronary Cusp

  • Author/Authors

    Yoshiyuki Takami، نويسنده , , Hiroshi Masumoto، نويسنده , , Biliie Fyfe-Kirschner، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    2210
  • To page
    2213
  • Abstract
    We present a case of disruption of the porcine aortic wall of the 27-mm Freestyle stentless bioprosthesis 5 years after the subcoronary implantation to exclude the sinus of Valsalva aneurysm of the noncoronary cusp. At the urgent reoperation, the inflow suture line was found to be intact, and therefore a new stented valve was sutured with the inflow Dacron cuff after removal of ruptured valve. The subcoronary implantation technique creates a cavity between the prosthetic and native aortic walls filled with hematoma. The outflow suture line dehiscence caused blood flow into the cavity, porcine aortic wall rupture, and leaflet destruction.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    2007
  • Journal title
    The Annals of Thoracic Surgery
  • Record number

    610752