• Title of article

    Off-Pump Coronary Artery Bypass Graft Surgery With a Pulsatile Catheter Pump for Left Ventricular Dysfunction

  • Author/Authors

    Massimo A. Mariani، نويسنده , , Jan C. Diephuis، نويسنده , , Martin J.H. Kuipers، نويسنده , , Monica Gianoli، نويسنده , , Jan G. Grandjean، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    3
  • From page
    690
  • To page
    692
  • Abstract
    We describe the use of a novel device, the pulsatile catheter pump, in patients with left ventricular dysfunction undergoing off-pump coronary surgery. During a 1-year period, 14 patients (mean ejection fraction 28% ± 8%) underwent off-pump coronary surgery using the pulsatile catheter pump. We recorded neither mortality nor major adverse cardiovascular and cerebral events. Mean support time was 55 ± 13 minutes. The average flow generated by the pulsatile catheter pump, as calculated per patient, was 2.4 ± 0.2 L/min (range, 2.2 to 2.8 L/min). Our results show that the pulsatile catheter pump is clinically safe and provides adequate mechanical circulatory support in patients with impaired left ventricular function undergoing off-pump coronary artery surgery.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    2007
  • Journal title
    The Annals of Thoracic Surgery
  • Record number

    610928