Title of article
Alternative Technique for Implanting an Implantable Cardioverter Defibrillator in Infants
Author/Authors
Shoji Suzuki and Shigeru Sato، نويسنده , , Hiroshi Watanabe، نويسنده , , Shinpei Yoshii، نويسنده , , Shigeaki Kaga، نويسنده , , Yoshihiro Honda، نويسنده , , Natsuya Ishikawa، نويسنده , , Masahiko Matsumoto، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
3
From page
1701
To page
1703
Abstract
We applied a new implanting technique for an implantable cardioverter defibrillator (ICD) in a 4-month-old girl with repeated ventricular fibrillation (Vf) due to long QT syndrome. This technique consisted of placement of an oval patch lead on the outer surface of the pericardium in the left pleural cavity. This was useful in preventing the complications of the conventional epicardial patch leads (ie, crinkling of the lead and constrictive pericarditis). This patch should be contemplated as an alternative option for implanting ICD in infants.
Journal title
The Annals of Thoracic Surgery
Serial Year
2008