• Title of article

    A novel method to reduce device-related infections in patients supported with the HeartMate device

  • Author/Authors

    Latif Arusoglu، نويسنده , , Reiner Koerfer، نويسنده , , Gero Tenderich، نويسنده , , William A. Alexander، نويسنده , , Aly El-Banayosy، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1999
  • Pages
    3
  • From page
    1875
  • To page
    1877
  • Abstract
    Improved implantable left ventricular assist device technology has made survival to heart transplantation a near certainty. Nevertheless, infection remains a major risk to recipients of current percutaneous systems. We developed a modified implantation technique applied to the last 9 of 30 patients who received the HeartMate vented electric left ventricular assist system (LVAS). Covering the upper surface of the pump with a patch of knitted graft material was followed by a decline in the incidence of pocket infections from 33.3% to 11.1%. This modification compares favorably to that of a lengthened percutaneous driveline tunnel in reducing device-related infection.
  • Journal title
    The Annals of Thoracic Surgery
  • Serial Year
    1999
  • Journal title
    The Annals of Thoracic Surgery
  • Record number

    616346