• Title of article

    Thermal Injury Caued by Chandelier Fiber Probe

  • Author/Authors

    Hiroyuki himada، نويسنده , , Hiroyuki Nakahizuka، نويسنده , , Takayuki Hattori، نويسنده , , Ryuaburo Mori، نويسنده , , Yohihiro Mizutani، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    3
  • From page
    167
  • To page
    169
  • Abstract
    Purpoe To invetigate thermal change produced by a chandelier fiber probe (CFP). Deign Interventional ingle cae report and laboratory invetigation. Method When CFP wa ued for 25-gauge vitrectomy, an abnormal odor wa emitted, and the CFP wa intantly removed. To invetigate thi incident, the thermal change produced by the CFP wa meaured in air by mean of an infrared enor. Porcine Tenon capular tiue wa ued to reproduce the intraoperative phenomenon. Reult In our patient, the CFP tip melted, but thi did not affect viual function. In the experiment, the uncoated CFP reached 49 C (120 F) at maximum illumination. When porcine Tenon capular tiue wa attached to the CFP tip, a temperature of 144 C (291 F) wa reached at 50% of maximum illumination, and over 201 C (394 F) at 75%. The tip melted under thee condition. Concluion At preent, we avoid the ue of CFP during air exchange. Thi potential hazard of the CFP need to be addreed.
  • Journal title
    American Journal of Ophthalmology
  • Serial Year
    2007
  • Journal title
    American Journal of Ophthalmology
  • Record number

    626702