Title of article
Lack of agreement between tympanic and oral temperature measurements in adult hospitalized patients, ,
Author/Authors
Farrin A. Manian، نويسنده , , Sandy Griesenauer، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
3
From page
428
To page
430
Abstract
Background: The purpose of this study was to compare temperature measurements obtained by tympanic thermometers with those obtained by oral electronic or mercury-glass thermometers in adult hospitalized patients.
Methods: A prospective study of 406 nonintensive care unit adult patients hospitalized during an 8-month period in a tertiary care community medical center.
Results:Poor agreement was observed between tympanic versus electronic thermometer readings, with 95% limits of agreement of –2.11°F to +2.81°F. Similarly, poor agreement was observed between tympanic versus oral mercury-glass temperatures, with 95% limits of agreement of –1.72°F and +2.64°F. Of 27 oral electronic temperatures 100°F (37.7°C) or higher, 10 (37%, 95% confidence interval 19% to 58%) readings were 99.5°F (37.5°C) or lower, and six (22%, 95% confidence interval 9% to 42%) measured lower than 98.6°F by tympanic thermometers.
Conclusions: Temperatures measured by tympanic thermometers generally have poor agreement with those measured by oral electronic or mercury-glass thermometers in adult hospitalized patients. We recommend that tympanic thermometers not be used for routine screening for fever in this patient population.
Journal title
American Journal of Infection Control (AJIC)
Serial Year
1998
Journal title
American Journal of Infection Control (AJIC)
Record number
635137
Link To Document