• Title of article

    Flow Injection Potentiometric Determination of Cd2+ Ions Using a Coated Graphite Plasticized PVC-Membrane Electrode Based on 1,3-Bis(2-cyanobenzene)triazene

  • Author/Authors

    Mojtaba Shamsipur، نويسنده , , Shokat Sahari، نويسنده , , Mahmood Payehghadr، نويسنده , , Kamal Alizadeh، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    8
  • From page
    555
  • To page
    562
  • Abstract
    1,3-Bis(2-cyanobenzene)triazene, L, was used as a suitable ionophore for the fabrication of a new PVC-based polymeric membrane coated graphite electrode for selective sensing of Cd2+ ion. The electrode exhibited a selective linear Nernstian response to Cd2+ ion at an optimal pH range of 6-9 with a limit of detection of 8.0 × 10–6 M and a fast response time of about 2 s. The electrode was used as a proper detection system in flow-injection potentiometry of cadmium ion and resulted in well defined peaks for cadmium ions with stable baseline, excellent reproducibility and high sampling rates of over 100 injections per hour. It showed good stability, reproducibility and fast response time. The practical utility of the proposed system has also been reported.
  • Keywords
    Coated graphite electrode , 1 , 3-bis(2-cyanobenzene)triazene , cadmium ion determination , Flow injection potentiometry
  • Journal title
    Acta Chimica Slovenica
  • Serial Year
    2011
  • Journal title
    Acta Chimica Slovenica
  • Record number