• Title of article

    Tendon sheath analysis for estimation of distal end force and elongation for sensorless distal end

  • Author/Authors

    S. J. Phee، نويسنده , , S. C. Low، نويسنده , , P. Dario and A. Menciassi، نويسنده ,

  • Issue Information
    دوماهنامه با شماره پیاپی سال 2010
  • Pages
    10
  • From page
    1073
  • To page
    1082
  • Abstract
    Tendon sheath actuation is found in many applications, particularly in robotic hands and surgical robots. Due to the friction between the tendon and sheath, many undesirable characteristic such as backlash, hysteresis and non-linearity are present. It is desirable to know the end-effector force and elongation of the tendon to control the system effectively, but it is not always feasible to fix sensors at the end effector. A method to estimate the end-effector parameters using only a force and position sensor at the proximal site is given. An analytical study is presented and experiments are reported to support the result, showing a maximum full-scale error of approximately 7%. This result is achieved if the shape of the sheath remains the same and buckling is negligible. The results presented in this study could contribute towards haptic development in robotics surgery.
  • Keywords
    Grasping , Surgical Robots , force control , Serial manipulator design and kinematics , Haptic interfaces
  • Journal title
    Robotica
  • Serial Year
    2010
  • Journal title
    Robotica
  • Record number