Title of article
Differential Pulse Voltammetric Assay of Antibacterial Drug Doxycycline Hyclate
Author/Authors
Attia، Hazem Ali نويسنده , , Saber ، Refaat A. نويسنده National Organization for Drug Control and Research, P.O. Box 29, Cairo, Egypt ,
Issue Information
دوماهنامه با شماره پیاپی سال 2011
Pages
11
From page
291
To page
301
Abstract
The electrochemical oxidation of Doxycycline Hyclate has been carried out in
Britton - Robinson buffer at carbon paste and glassy carbon electrodes. Doxycycline Hyclate
exhibits a well-defined irreversible oxidation peak in a broad pH range (2-11). Differential
pulse voltammetry was used to determine Doxycycline Hyclate in pure form. The peak
current varied linearly in the following ranges: 2.0×10-7- 3.0×10-6 mol L-1 and 2.0×10-7
- 2.6×10-6 mol L-1 in case of carbon paste electrode and glassy carbon electrode, respectively.
In case of carbon paste electrode the limits of detection (LOD) and quantification (LOQ) were
6.56×10-8 mol L-1 and 2.19×10-7 mol L-1, respectively. For glassy carbon electrode the LOD
and LOQ were 9.55×10-8 mol L-1 and 3.18×10-7 mol L-1, respectively. The percentage
recoveries were found in the following ranges: 99.58-101.16% and 99.46-100.98% for carbon
paste electrode and glassy carbon electrode, respectively. The relative standard deviations
were found in the following ranges: 0.53-1.79% and 0.48-1.96% in case of carbon paste
electrode and glassy carbon electrode, respectively. Differential pulse voltammetry was
successfully applied for the determination of Doxycycline Hyclate in pharmaceutical form.
Journal title
Analytical and Bioanalytical Electrochemistry
Serial Year
2011
Journal title
Analytical and Bioanalytical Electrochemistry
Record number
691434