Title of article
Electrical calibration method for cascade impactors
Author/Authors
Jorma Keskinen، نويسنده , , Marko Marjam?ki، نويسنده , , Annele Virtanen، نويسنده , , Timo M?kel?، نويسنده , , Risto Hillamo، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 1999
Pages
6
From page
111
To page
116
Abstract
A new method for measuring collection characteristics of cascade impactors is described. The method is based on challenging the impactor with monodispersed charged particles and measuring electrical currents on all the impactor stages plus a backup filter. From the current values, collection efficiency at a certain particle diameter can be calculated simultaneously for each impactor stage. To test the method, monodispersed dioctyl sebacate particles were generated using an atomiser, evaporation/condensation generator and an electrostatic classifier. The particle collection characteristics of the six lowest stages of a Berner-type low-pressure cascade impactor were measured. Comparison measurements were made with a real-time stage efficiency measurement system for single impactor stages. The results of the two methods were in good agreement.
Journal title
Journal of Aerosol Science
Serial Year
1999
Journal title
Journal of Aerosol Science
Record number
742500
Link To Document