Title of article
Characterisation of a unipolar charger for droplet aerosols of 0.1–20 μm in diameter
Author/Authors
G?ran P. Frank، نويسنده , , Sven?Inge Cederfelt، نويسنده , , Bengt G. Martinsson، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2004
Pages
18
From page
117
To page
134
Abstract
A charging unit for unipolar charging of cloud and fog droplets has been developed and calibrated regarding ion current, angular distribution of ions, mean droplet charge and droplet charge distribution. The charger is part of the droplet aerosol analyser, which is an instrument used for characterisation of droplets and interstitial particles in atmospheric clouds and fog. Three different calibration methods have been used for the charge calibration, whereof the third and newest method makes use of a differential mobility analyser, especially developed for large particles, up to at least 20 μm in diameter.
The number of charges acquired is a strictly increasing function of the droplet diameter. Two charging models have been used for comparisons with the measurements and both models show good agreement with the data regarding the size dependency of the charge level. Both models can thus be used to describe the performance of the charger.
Keywords
Droplet Aerosol Analyser , DAA , Cloud droplet , Fog droplet , Unipolar charging