• Title of article

    Intercalibration of SUMER and CDS on SOHO. III. SUMER and CDS-GIS

  • Author/Authors

    Pauluhn، Anuschka نويسنده , , Lang، Jim نويسنده , , Breeveld، Eddie R. نويسنده , , Solanki، Sami K. نويسنده , , Schühle، Udo نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -656
  • From page
    657
  • To page
    0
  • Abstract
    A collinear photothermal detection bench is described that makes use of a position-modulated heating source instead of the classic power-modulated source. This new modulation scheme increases by almost a factor 2 the sensitivity of a standard mirage bench. This bench is then used to measure the absorption coefficient of OH-free synthetic fused silica at 1064 nm in the parts per 106 range, which, combined with spectrophotometric measurements, confirms that the dominant absorption source is the OH content. [Optical Society of America ]
  • Keywords
    instrumentation , Measurement , Metrology , Radiometry , Spectrometers , spectroscopic instrumentation , Spectroscopy , Ultraviolet
  • Journal title
    Applied Optics
  • Serial Year
    2003
  • Journal title
    Applied Optics
  • Record number

    74578