• Title of article

    Establishing a high-accuracy spectral response scale in the near infrared with digital filters

  • Author/Authors

    Gran، Jarle نويسنده , , Ellingsberg، Kristian نويسنده , , Sudbo، Aasmund S. نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    -2481
  • From page
    2482
  • To page
    0
  • Abstract
    Spectrally invariant detectors are commonly used to interpolate or extrapolate the responsivity of InGaAs detectors in the infrared from absolute calibrations at a few wavelengths. The random noise in such detectors limits the accuracy that can be achieved in a narrowband, double-monochromator setup. We propose the application of a dedicated digital filter, which reduces the uncertainty by 30%, and combine it by calibrating a group of three detectors. The uncertainties are propagated from the observed variance in the relative measurement to the combined uncertainty of 0.4% (2 sigma) in the responsivity values of the InGaAs detectors in the range of 1010-1640 nm.
  • Keywords
    Infrared , Photodetectors , Radiometry , Metrology
  • Journal title
    Applied Optics
  • Serial Year
    2005
  • Journal title
    Applied Optics
  • Record number

    75736