Title of article
Measurement of wave-front aberration in soft contact lenses by use of a Shack-Hartmann wave-front sensor
Author/Authors
Jeong، Tae Moon نويسنده , , Menon، Manoj نويسنده , , Yoon، Geunyoung نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
-4522
From page
4523
To page
0
Abstract
Lower- and higher-order wave-front aberrations of soft contact lenses were accurately measured with a Shack-Hartmann wave-front sensor. The soft contact lenses were placed in a wet cell filled with lens solution to prevent surface deformation and desiccation during measurements. Aberration measurements of conventional toric and multifocal soft contact lenses and a customized soft contact lens have proved that this method is reliable. A Shack-Hartmann wave-front sensor can be used to assess optical quality of both conventional and customized soft contact lenses and to assist in enhancing lens quality control.
Keywords
Medical optics instrumentation , Ophthalmic optics
Journal title
Applied Optics
Serial Year
2005
Journal title
Applied Optics
Record number
76195
Link To Document