    This study was designed to assess the pCO2 accuracy of portable mainstream (Tidal Wave, Novametrix; Propaq 106, Protocol) and sidestream capnometers (Capnocheck 8200, BCI; Capnocount mini, Weinmann; NPB-75, Nellcor Puritan Bennett; SC-210, Pryon) with respect to international standards and preclinical emergency conditions. Measurements were performed under temperature conditions of +22°C and −20°C using dry gas mixtures with different CO2 concentrations (STPD) and in patients ventilated with pure oxygen (BTPS). Accuracy presented to be between +1% (Capnocheck) and +12% (Propaq) (STPD) and between −0.4% (Capnocheck) and +11% (Tidal Wave) (BTPS). The measurements were affected by low ambient temperature only in the NPB-75 (+15%). Our results indicate that portable quantitative capnometers are able to fulfill accuracy requirements as requested by international standards but can be affected by changing ambient temperatures.
  • Keywords
    Capnometry , Capnography , Handheld , accuracy , Prehospital , cold ambient temperature , Emergency medicine , Out-of-hospital , Portable