Title of article
Effective electrode configuration for selective stimulation with inner eye prostheses
Author/Authors
F.، Rattay, نويسنده , , S.، Resatz, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
-1658
From page
1659
To page
0
Abstract
The quality of visual perception with retinal prostheses strongly depends on the local selectivity. Electrode arrays at the surface of the retina should excite exclusively cells within a local area but they are expected to co-stimulate bypassing axons originating from ganglion cells of the outer regions. Long electrodes parallel to these axons are shown to be good candidates for avoiding the co-stimulation phenomenon. Efficiency of focal excitation depends on the length and resistance of the electrodes.
Journal title
IEEE Transactions on Biomedical Engineering
Serial Year
2004
Journal title
IEEE Transactions on Biomedical Engineering
Record number
80545