Title of article
Large area gamma-ray imaging detector based on high-resolution CdTe diode
Author/Authors
H.، Nakamura, نويسنده , , K.، Nakazawa, نويسنده , , T.، Takahashi, نويسنده , , S.، Watanabe, نويسنده , , T.، Mitani, نويسنده , , Uno، Shingo نويسنده , , H.، Tajima, نويسنده , , M.، Nomachi, نويسنده , , Y.، Fukazawa, نويسنده , , S.، Kubo, نويسنده , , Y.، Kuroda, نويسنده , , M.، Onishi, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-1047
From page
1048
To page
0
Abstract
We are developing a large array detector composed of 1024 individual CdTe diodes. Each detector has the dimensions of 1.2 mm * 5.0 mm and a thickness of 1.2 mm. An edge-on geometry is used for the injection of (gamma)-rays, to obtain a crosssection thickness of 5 mm. With this geometry, the distance between the two electrodes can be kept small, and we can therefore apply the high electric field which is necessary to achieve a high energy resolution (by reducing the low energy tail) and also to sustain the long-term stability of the CdTe diode. Signals from each detector element are fed into newly developed low noise ASICs. We use 32 chips for the readout of 1024 elements. In this paper, we report the basic characteristics of the individual detectors and the overall performance of the gamma-camera. Design of the readout electronics system is also described.
Keywords
Laminated waveguide , low-temperature co-fired ceramic (LTCC) , millimeter wave , rectangular waveguide (RWG) , waveguide transition
Journal title
IEEE Transactions on Nuclear Science
Serial Year
2003
Journal title
IEEE Transactions on Nuclear Science
Record number
86458
Link To Document