• Title of article

    Comparison of independent optical frequency measurements using a portable iodine-stabilized Nd:YAG laser

  • Author/Authors

    Ye، Jun نويسنده , , Hong، Feng-Lei نويسنده , , H.، Matsumoto, نويسنده , , J.L.، Hall, نويسنده , , J.، Ishikawa, نويسنده , , K.، Sugiyama, نويسنده , , A.، Onae, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -23
  • From page
    24
  • To page
    0
  • Abstract
    A comparison of two independent absolute optical frequency measurements has been carried out between the National Metrology Institute of Japan/National Institute of Advanced Industrial Science and Technology (NMIJ/AIST), Tsukuba, Japan and the JILA (formerly the Joint Institute for Laboratory Astrophysics), Boulder, CO, using a portable iodine-stabilized Nd:YAG laser. The agreement between the two absolute measurements is better than the measurement uncertainty of 6.7*10^-13 that can be attributed to the reproducibility limitations of the portable laser. This comparison is used to confirm the measured absolute frequency of an iodine-stabilized Nd:YAG laser at NMIJ/AIST (Y3), which is reported to the Consultative Committee for Length (CCL) for the determination of the absolute frequency value of iodine-stabilized Nd:YAG lasers.
  • Keywords
    leukemia
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Serial Year
    2003
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Record number

    91455