Title of article
Two ions in one trap: ultra-high precision mass spectrometry?
Author/Authors
S.، Rainville, نويسنده , , J.K.، Thompson, نويسنده , , D.E.، Pritchard, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-291
From page
292
To page
0
Abstract
By trapping two different ions in the same Penning trap at the same time, we simultaneously measured the ratio of their cyclotron frequencies (from which we obtain their atomic mass ratio) with a precision of about 10^-11 in only a few hours. In order to perform these comparisons, we must be able to measure and control all three normal modes of motion of each ion - cyclotron, axial, and magnetron. We have developed novel techniques to do, so and we are currently using these tools to carefully investigate the important question of systematic errors in those measurements. This new technique shows promise to expand the precision of mass spectrometry an order of magnitude beyond the current state-of-the-art.
Keywords
leukemia
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year
2003
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number
91467
Link To Document