• Title of article

    Projecting Zener DC reference performance between calibrations

  • Author/Authors

    C.A.، Hamilton, نويسنده , , L.W.، Tarr, نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    -453
  • From page
    454
  • To page
    0
  • Abstract
    Two simple empirical models are proposed to predict Zener reference standard output voltage and uncertainty based on just one or two calibration points. The models are derived from a study of 140 Zener standards. The results yield a substantial reduction in uncertainty relative to the manufacturerʹs specification and illustrate the dominant influence of seasonal effects on projection uncertainty.
  • Keywords
    leukemia
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Serial Year
    2003
  • Journal title
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
  • Record number

    91505