Title of article
Projecting Zener DC reference performance between calibrations
Author/Authors
C.A.، Hamilton, نويسنده , , L.W.، Tarr, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-453
From page
454
To page
0
Abstract
Two simple empirical models are proposed to predict Zener reference standard output voltage and uncertainty based on just one or two calibration points. The models are derived from a study of 140 Zener standards. The results yield a substantial reduction in uncertainty relative to the manufacturerʹs specification and illustrate the dominant influence of seasonal effects on projection uncertainty.
Keywords
leukemia
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year
2003
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number
91505
Link To Document