Title of article
Effects of metallic gates on AC measurements of the quantum Hall resistance
Author/Authors
B.، Jeckelmann, نويسنده , , B.، Jeanneret, نويسنده , , F.، Overney, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-573
From page
574
To page
0
Abstract
Using a sample with a split back gate, a linear frequency dependence of the AC quantum Hall resistance (QHR) was observed. The frequency coefficient, which is due to dielectric losses produced by leakage current between the two-dimensional electron gas (2DEG) and the back gates, can be turned from a positive to a negative value by increasing the back-gate voltage. More interestingly, by removing these back gates, the losses can be considerably reduced, leading to a residual frequency coefficient on the order of (0.03 +- 0.03) . 10^-6 /kHz. Moreover, at 1 kHz, an extremely flat plateau was observed over a magnetic field range of 1.4 T. These results clearly indicate that the audio frequency dependence of the QHR is to a large extent related to the measurement apparatus and does not originate from the physical transport properties of the 2DEG.
Keywords
leukemia
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year
2003
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number
91533
Link To Document