Title of article
NIST-NPL interlaboratory pulse measurement comparison
Author/Authors
N.G.، Paulter, نويسنده , , D.R.، Larson, نويسنده , , A.J.A.، Smith, نويسنده , , T.M.، Souders, نويسنده , , A.G.، Roddie, نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
-1824
From page
1825
To page
0
Abstract
A comparison of the pulse parameter values obtained from the pulse measurement services of the National Institute of Standards and Technology, USA, and the National Physical Laboratory, U.K., was performed. The comparison was based on the pulse parameters of amplitude, transition duration, overshoot, and undershoot (preshoot). The parameter comparison was applied to raw (measured) waveforms, corrected waveforms (if applicable), and reconstructed waveforms. The results of the comparison show that the pulse parameter values for both national laboratories are within published uncertainties.
Keywords
leukemia
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Serial Year
2003
Journal title
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
Record number
91713
Link To Document