Author/Authors
Mehmet GULfiEN، نويسنده , , Vasfi KARATOSUN، نويسنده , , Bahadir UYULGAN، نويسنده ,
DocumentNumber
1952587
Title Of Article
The biomechanical assessment of fixation methods in periprosthetic femur fractures
شماره ركورد
12576
Latin Abstract
Objective: The aim of our study was to compare the biomechanical properties of different fixa¬tion methods used in periprosthetic femur fractures. Methods: We created sawbone models with Mallory Type 2 periprosthetic femur fractures. The periprosthetic fractures were fixed with cables, cables and strut graft, or cables and plates. The biomechanical properties of these three different fixation methods were compared with the intact femur, the intact femur with prosthesis and the femur with periprosthetic fracture without fixa¬tion. Results: The periprosthetic fracture without fixation had a significantly lower yield point value than the periprosthetic fractures with fixation (p<0.05). There was no significant difference between the three different fixation methods of the periprosthetic fractures. The intact femur with a prosthesis showed statistically higher values than all three fixation methods of perioprosthetic fractures. Conclusion: Mallory Type 2 periprosthetic fractures should be fixed. There is no difference among the fixation methods used in the study and none of them provide a stability equivalent to that of an intact femur with prosthesis.
From Page
266
NaturalLanguageKeyword
Arthroplasty , biomechanics , periprosthetic , Composite , Fracture , Hip , Femur
JournalTitle
Studia Iranica
To Page
269
To Page
269
Link To Document