Author/Authors
Duman, Ramazan Kriminal Polis Laboratuvarı, Turkey , Doğan, Selda Mersin Üniversitesi - Eczacilik Fakültesi, Turkey , Kaya, F. Nazlı Dinçer Mersin Üniversitesi - Eczacılık Fakültesi, Turkey
Title Of Article
Investigation of Effect of Nitric Acid and Various Matrix Modifier on Determination of Lead in Biological Samples by Electrothermal Atomic Absorption Spectroscopy
شماره ركورد
35961
Abstract
In this study, determination of lead in biological materials by electrothermal atomic absorption spectroscopy was investigated. Pyrolysis and atomization temperatures were optimized in Pt + Pd, Pd + Mg(NO3)2 and NH4H2PO4 + Mg(NO3)2 mixtures using various nitric acid concentrations. % 10 HNO3 + NH4H2PO4 + Mg(NO3)2 mixtures was designated as the best conditions. The accuracy of the method was tested by certificated reference standard materials, CRM BCR-397 human hair, NIST 2976 mussel tissue and HIGH EP-H-1 drinking water. Relative errors were less than 10 % for all of the certificated reference standard materials. Limit of detection (LOD) and limit of quantification (LOQ) values were calculated as 0,67 μg/L and 2,01.
From Page
23
NaturalLanguageKeyword
Lead determination , matrix modifier , electrothermal atomic absorption spectroscopy , LOQ
JournalTitle
Süleyman Demirel University Faculty of Arts and Science Journal of Science
To Page
33
JournalTitle
Süleyman Demirel University Faculty of Arts and Science Journal of Science