DocumentCode
1011271
Title
Hydraulic resistance and damping in catheter-transducer systems
Author
Taylor, Bruce C. ; Donovan, Francis M., Jr.
Author_Institution
Dept. of Biomed. Eng., Akron Univ., OH, USA
Volume
11
Issue
4
fYear
1992
Firstpage
72
Lastpage
78
Abstract
Research undertaken to investigate the characteristics of typical catheter-transducer systems to determine how frequency-dependent resistance and damping values can affect the reproductive fidelity of direct arterial pressure recordings is discussed. While the data presented will not cause methods and techniques used in invasive pressure monitoring to be altered appreciably, they will explain some of the anomalies reported in predicting system damping from measurable system parameters, and may aid in the selection of optimal components for accurate recording of blood pressure.<>
Keywords
biomedical equipment; damping; haemodynamics; pressure measurement; transducers; blood pressure recording; catheter-transducer systems; damping; direct arterial pressure recordings; frequency-dependent resistance; hydraulic resistance; invasive pressure monitoring; Blood pressure; Catheters; Damping; Electric resistance; Heart rate; Immune system; Navier-Stokes equations; Pressure measurement; Steady-state; Viscosity;
fLanguage
English
Journal_Title
Engineering in Medicine and Biology Magazine, IEEE
Publisher
ieee
ISSN
0739-5175
Type
jour
DOI
10.1109/51.256965
Filename
256965
Link To Document