• DocumentCode
    1011271
  • Title

    Hydraulic resistance and damping in catheter-transducer systems

  • Author

    Taylor, Bruce C. ; Donovan, Francis M., Jr.

  • Author_Institution
    Dept. of Biomed. Eng., Akron Univ., OH, USA
  • Volume
    11
  • Issue
    4
  • fYear
    1992
  • Firstpage
    72
  • Lastpage
    78
  • Abstract
    Research undertaken to investigate the characteristics of typical catheter-transducer systems to determine how frequency-dependent resistance and damping values can affect the reproductive fidelity of direct arterial pressure recordings is discussed. While the data presented will not cause methods and techniques used in invasive pressure monitoring to be altered appreciably, they will explain some of the anomalies reported in predicting system damping from measurable system parameters, and may aid in the selection of optimal components for accurate recording of blood pressure.<>
  • Keywords
    biomedical equipment; damping; haemodynamics; pressure measurement; transducers; blood pressure recording; catheter-transducer systems; damping; direct arterial pressure recordings; frequency-dependent resistance; hydraulic resistance; invasive pressure monitoring; Blood pressure; Catheters; Damping; Electric resistance; Heart rate; Immune system; Navier-Stokes equations; Pressure measurement; Steady-state; Viscosity;
  • fLanguage
    English
  • Journal_Title
    Engineering in Medicine and Biology Magazine, IEEE
  • Publisher
    ieee
  • ISSN
    0739-5175
  • Type

    jour

  • DOI
    10.1109/51.256965
  • Filename
    256965